Scanner Settings
Scientific scanners and imagers are complex instruments. The
following
section should give you some background information to make the right
choices
for scanning your gels.
The principle of operation of a scanner is quite simple: a pixel in
the image
file is created from light intensity that is measured at a certain
point in the
gel. In more detail, light is emitted from a light source, reflected or
transmitted by the gel,
and then measured by a detector. The measured value is converted to a
number by
an A/D (analog to digital) converter. The numbers from the whole gel
are then
processed to give the image file that is the output of the scanning
process.
Choosing Scan Mode
The Scan Mode describes how the light travels from the light source
to the
detector.
- In transmisson mode the light passes through the gel and
is directly
measured by the detector on the opposite side of the gel (see left
figure
above).
- In reflection mode the gel is between the light source
and a
reflector (normally in the scanners lid). Light travels through the
gel, is
reflected, goes again through the gel, and is then detected.
Three scanning principles: light transmission, light
reflection, fluorescence
 |
For high quality data we recommend using transmission mode because
saturation effects are less likely than with reflection mode.
Saturation means
that differences in higher spot intensities are not properly resolved
-- dark
spot centers become completely black. In reflection mode this happens
more
often because the light travels twice through the gel, so it can be
more
easily absorbed completely. Saturated regions appear as "plateau spots"
in the
3D view of a spot (see the following figure).
Saturated "plateau spot"
 |
Consideration of the scan mode is important if you use staining
techniques
absorbing the light, e.g. all types of Coomassie or silver nitrate
staining. For
fluorescence scanning the scan mode is less relevant because the light
is
emitted by the proteins within the gel.
Light intensity
The intensity of the light source directly influences
differentiation of high
density spots. It is common to use office document scanners to detect
absorbing
dyes (Coomassie, Silver, x-ray film). The manufacturers usually claim a
differentiation of gray levels until an optical density (OD) of 2.0.
This is
mostly sufficient for gel documentation purposes. However, it is
insufficient
for differentiation of high density spots reaching optical densities
near 3.7 to
4.0.
Document scanners resolving ODs from 3.7 to 4.0 can be found in the
premium
or professional product lines of well known manufacturers (e.g.
Quato, Umax).
Warming up the lamp before starting the scan process is
highly
important because the lamps become more powerful when they reach their
working
temperature. Extensive usage of the scanner causes aging of the light source
resulting in weaker light intensity.
Especially for fluorescence imaging high performance light
sources
are absolutely necessary. In this field of imaging normally lasers or
UV lamps
are used. Lasers deliver highly concentrated light and can effectively
excite
fluorescent dyes. Depending on the fluorescent dyes different colored
lasers
and excitation and emission filters are needed. Lasers need a
warm up
phase too to reach their highest performance.
The light harvesting detector and the A/D converter collect
the
light that has passed through the gel and translate this information to
computer readable signals. The detector (CCD element or PMT) converts
the light
intensity to an electrical signal that is subsequently digitized by the
A/D
(analog to digital) converter. These data are stored in the image file.
In many high perfomance imaging devices for fluorescence detection
also very
weak signals have to be detected. That is why these scanners allow
to tune the
voltage of the light detector (PMT - photo multiplier tube). Higher
voltages
increase the sensitivity for very weak signals but also increase image
background. Unfortunately increasing PMT voltage also results in
amplifying
noise. Because of this fact PMT voltage should be modified with care.
Another
pitfall is simultanous enhancement of strong signals bringing very
intense
spots to saturation.
In desktop scanners the A/D converter very often can differentiate a
larger
dynamic range of intensities than what can be stored in the file
format. This
is the reason, why the internal (A/D converter) color depth (e.g. 12
bit) may
be larger than the output (e.g. 8 bit). The advantage of such a
constellation
is the flexibility for the selection of the right intensity range after
a
prescan for the final data storage. An alternative approach is the data
processing of the complete data coming from the A/D converter by using
gradation curves (see also image calibration).
Intensity range
Enlarging the intensity range between signal and background
is one of
the major tasks in image generation. The larger the distance between
background
and top intensity on an image the larger the dynamic range that is
available
for coding intensity levels. The more intensity levels are coded in the
image
the more accurate is the quantitative analysis. There are several ways
to
increase the signal to background range:
- Cutting off a high background by a scanner preset. In the prescan
the user can
traditionally specify the lowest and highest intensity signals that
will be
used as lowest and highest intensity signal in the resulting image
file. The
scanner will then distribute the measured intensity values evenly
between the
given minimum and maximum. This makes sense if the A/D converter can
distinguish
more intensity levels than what can be stored in the image file. Note
that this
approach only reaches highest performance if you avoid scanning areas
that are
not covered by the gel.
- Controlling the conversion from color to grayscale. Most office
scanners
create grayscale images by converting a color scan to a grayscale image
file.
This is done by averaging intensities over the different color
channels. You
can improve the resulting image in some situations by controlling the
color to
grayscale conversion yourself. For example:
- A coomassie blue stained image can be better scanned by
excluding the blue
color channel from the color image befor converting it to a grayscale
image.
- A similar approach can be used for the brownish silver
stains. As a rule,
extract those color channels with high absorption. Brownish silver
stains
absorb the blue light most efficiently, so extracting the blue color
channel
will give the best results.
- Increasing the voltage of the light detector. If the intensity
range is not
completely used by intensity signals in a fluorescence scan the voltage
of the
light detector may be increased. This will increase the signals overall
but
also increases the absolute distance between background signal and
strongest
intensity signal on the gel image. Please be aware of saturation
effects!
A perfect spot without any saturation effects
 |
Saturation effects heavily affect quantitative data because
they
virtually truncate the peaks of high intensity spots. These truncations
distort
the quantities of spots. Many scanner software packages highlight
saturated
image regions. Saturation effects may occur for several reasons:
Intense staining can cause complete absorbtion of light passing through the spot. This
results in planar truncated spots in the 3D view of the gel image. Try to use a
scanner with a more intense light source (e.g. a laser scanner) or to warm up
the scanner's lamp for a longer time. Reducing staining or increasing destaining time
may also solve this problem.
Dyes can appear in high concentration. Some staining techniques
accumulate pigment on top of high intensity spots losing the amount /
intensity correlation. This can be observed in all techniques based on
the
photographic silver nitrate staining (silver staining, x-ray film).
While for
classical silver staining techniques the typical doughnut spots may be
observed
the saturations from x-ray films have a planar appearance. To avoid
these
problems use shorter staining or developing times, or load less protein
on the
gel. Silver staining should not be used in quantitative proteome analysis
because
of saturation problems, and, more importantly, because the relation
between
silver stain intensity and protein amount is usually not linear.
A similar effect can be observed during phosphorimaging. If a gel is
overexposed to a phosphorscreen the screen's capacity to accumulate
energy from
radioactive decay events may be exhausted. This also results in planar
truncated spots in the 3D view. Here reducing the exposure time will
solve the
problem.
Heavily fluorescing gels may overload the light detector or heavily excited
light detectors may overload the A/D converter. This results in a loss of
differentiation ability of strong signals. This also results in planar truncated
spots. Here reduction of the PMT voltage may help.
Silver stain: A "doughnut protein spot" that lost it's central quantity.
 |
Subsequent manipulation of the image data by using image processing software
not designed for quantitative image analysis may ruin your images. Visually
enhanced images often show truncations in the high and low intensity regions.
Gel image files are not only pictures but also collections of quantitative
data. Only flipping, morroring, rotating in 90 degrees steps and cropping
leave the original data intact. Free rotating, enlarging or reducing, any kind
of gray level, contrast or gamma adjustion distorts the quantitative data and
may destroy your image files (see also image calibration).
|